GasAlertMax
User Manual
Pump Calibration
The pump needs recalibration if the pump alarm continues
for more than 5 seconds and the LCD displays that the inlet
is blocked.
Note
The pump can only be calibrated if the above
screen is shown.
1. Press and hold C for 3 seconds until the
audible alarm beeps.
2. Wait for the auto zero function to complete. The
LCD displays PumP rEcAl.
3. Then the audible alarm beeps twice, J remains lit,
M flashes, and the LCD displays the
following:
4. To complete the calibration, block the pump inlet
(within 8 seconds).
40